KENCO ENGINEERING COMPANY
P.O. BOX 470426 TULSA, OK 74147-0426 PHONE: (918) 663-4406 FAX: (918) 663-4480
www.kenco-eng.com e-mail: [email protected]
SMARTSONIC INTEGRAL ACOUSTIC WAVE LEVEL / OPEN CHANNEL FLOW TRANSMITTER
INSTALLATION / OPERATION INSTRUCTIONS
TABLE OF CONTENTS
1. GENERAL DESCRIPTION ............................................................. 2
2. PRINCIPLE OF OPERATION ......................................................... 2
3. MODEL DESCRIPTION .................................................................. 2
4. INSTALLATION .............................................................................. 2
5. SPECIFICATIONS .......................................................................... 2
6. DIMENSIONS.................................................................................. 3
7. MOUNTING GUIDELINES .............................................................. 3
8. MOUNTING INSTRUCTIONS ......................................................... 4
9. WIRING ........................................................................................... 4
Wiring Connections....................................................................... 5
Loop Resistance Table (Intrinsically Safe).................................. 5
Loop Resistance Table (Non-Intrinsically Safe).......................... 5
10. CALIBRATION................................................................................ 6
Display / Keypad............................................................................ 6
11. INITIAL SET-UP PROGRAM .......................................................... 7
Initial Set-up Program Menu......................................................... 7
Five Initial Programs (Pr.20 thru Pr.24) ....................................... 8
12. STANDARD PROGRAMS .............................................................. 11
Standard Program Menu............................................................... 11
Programs (Pr.01 thru Pr.10).......................................................... 12
13. OPEN CHANNEL FLOW MEASUREMENT ................................... 21
Program Pr.00 (Flow Measurement) ............................................ 21
Flume / Weir Styles and Dimensions ........................................... 22
Flumes / Weirs – U.S. Standard.................................................... 23
Flumes / Weirs – European........................................................... 26
14. TROUBLESHOOTING .................................................................... 29
15. APPENDIX A (Gas Factor Table) .................................................. 30
Przeglądanie stron 0
1 2 3 4 5 6 ... 29 30

Podsumowanie treści

Strona 1 - KENCO ENGINEERING COMPANY

Page 1 KENCO ENGINEERING COMPANY P.O. BOX 470426 TULSA, OK 74147-0426 ● PHONE: (918) 663-4406 FAX: (918) 663-4480 www.kenco-eng.com e-mail: info@ken

Strona 2

Page 10 Initial Set-up Program Pr.21 (Flow Measurement Units) This program is only available if you selected Flow in Program 24. This is where you se

Strona 3

Page 11 STANDARD PROGRAMS Standard Program Menu FlowMeasurementResetTankHeightInterferingSignals4mA20mAResponseSpeedWorkingAreaLevel / DistanceAtmosph

Strona 4

Page 12 Reset Program The Pr.01 program enables you to do a reset, refreshing the SmartSonic measurement reading, while not changing data saved

Strona 5

Page 13 Press / Action Display Description and simultaneously Enter the Program Menu One Time Cycle to Program #2 Enter P

Strona 6 - Next button

Page 14 Function 00.16 (22mA Error Signal) This function allows you to define if “Near Zone” and “Lost Echo” are allowable errors, and if the result i

Strona 7

Page 15 Each reading (scan distance) taken using the Pr.03 program is stored as an interfering signal, until a readin

Strona 8

Page 16 Temporary display while SmartSonic searches for Interfering Signals After a brief pause For Example: Distance to Interfering Signal Sav

Strona 9

Page 17 and For Example: Change the value to represent the desired 4mA Point in either: Level, Distance, or Flo

Strona 10

Page 18 and For Example: Change the value to represent the desired 2mA Point in either: Level, Distance, or Flo

Strona 11

Page 19 Working Area The Pr.07 program allows you to add a distance range that exceeds the tank’s height, thus enabling accurate readi

Strona 12

Page 2 GENERAL DESCRIPTION The SmartSonic Acoustic Wave Transmitter is designed for high accuracy level measurement of a variety of liquids. The Smar

Strona 13

Page 20 For Example: Displays the current distance/level mode setting and Used to cycle between Distance (d000) and Level (L000)

Strona 14

Page 21 Restoring the Default Settings The Pr.10 program allows you to clear programs Pr.00, Pr.02 – Pr.09, and reverts to Factory Sett

Strona 15

Page 22 Press / Action Display Description and simultaneously Enter the Program Menu One Time Cycle to Program #00 For E

Strona 16

Page 23 Flumes / Weirs – U.S. Standard (dimensions are in inches unless otherwise indicated) Type 1 (X) – Rectangular Suppressed Sharp-Crested Weir

Strona 17

Page 24 Type 4 (X) – V-notch (Triangular) Sharp-Crested Weir Type 5 (X) - Parshall Flume Type 6 (X) – Palmer-Bowlus Flume

Strona 18

Page 25 Type 7 (X) – H-Flume Type 8 (X) – Leopold-Lagco Flume Code (FF) Flume Size Measurement Point 01 6 1.96 02 9

Strona 19

Page 26 Flumes / Weirs – European (dimensions are in centimeters unless otherwise indicated) Type 1 (X) – Rectangular Suppressed Sharp-Crested Weir

Strona 20

Page 27 Type 4 (X) – V-notch (Triangular) Sharp-Crested Weir Type 5 (X) – Khafagi-Venturi Flume Type 6 (X) – Pa

Strona 21

Page 28 Type 7 (X) – Palmer-Bowlus Flume Type 8 (X) – H-Flume Type 9 (X) – Neyrpic-Venturi Flume / Long Base Weir

Strona 22

Page 29 TROUBLESHOOTING This section describes how to resolve problems that may occur when calibrating / operating the SmartSonic. If the following i

Strona 23

Page 3 DIMENSIONS MOUNTING GUIDELINES 1.46 4.25 2.55 2.16 1.00 7.63 3.75 0.40 35º 1.18 The transmitter must be mounted in such

Strona 24

Page 30 Appendix A. – Gas Factor Table Atmosphere (Gas) Factor Atmosphere (Gas) Factor Acetic Acid 0.62 Ethane 0.90 Acetone 0.63 Ethanol 0.71

Strona 25

Page 4 MOUNTING INSTRUCTIONS Precautions: • Ensure that the transmitter is mounted in an area that meets the stated temperature, pressure and technic

Strona 26

Page 5 Wiring Connections Before starting installation procedures in hazardous areas, insure that all power sources have been turned off and locked o

Strona 27

Page 6 CALIBRATION This section explains how to set up and calibrate the SmartSonic for accurate measurement monitoring. SmartSonic is supplied with

Strona 28

Page 7 NOTE: Within some programs, the digits in the display can be individually modified. This is indicated by a flashing digit. In this case the E

Strona 29

Page 8 Five Initial Programs While the Pr.02 Program is normally used to set the Tank Height, there is a special code that is used to enter this Initi

Strona 30

Page 9 Initial Set-up Program Pr.23 (Measurement Range) This program allows you to set the transmitter to either Standard or Short Range. The Standar